Zenith Leander Eckard Auto Flexion Gas Charge Cylinder for CERVICAL TILT
Tilt head Adjustment for Raising By Itself 1. Extend Breakaway Chest all the way up 2. Loosen Jam Nut 3. Turn Shaft Counter clockwise 1/4 Turn 4. Check for Lock If still slipping turn another 1/4 Round till it doesn't slip, but still locks and unlocks. 5. Tighten Lock Nut back down Tilt head Adjustment for Not Releasing 1. Loosen Jam Nut 2. Turn Shaft Clockwise 1/4 Turn 3. Check for Lock / Unlock . 4.If not unlocking turn another 1/4 Round till it still locks and unlocks , but doesn't slip. 5. Tighten Lock Nut back down
Zenith Leander Eckard Auto Flexion Gas Charge Cylinder for CERVICAL TILT, Leander Gas Charge Cylinder, Leander Gas Tilt Cylinder, Leander Tilt, Leander Gas Strut, Leander Gas Cylinder, Zenith Leander Eckard Auto Flexion Gas Cylinder, Zenith Leander Eckard Auto Flexion Cylinder
Warranty Information
This Zenith Leander Eckard Auto Flexion Gas Charge Cylinder for CERVICAL TILT is 100% guarantee to be free of defects for one year from date of purchase.