Visit Our Other Websites

Questions?          Feel Free To Call Us At 256-245-8390 . We Are Open 9-5 Central Time Zone.

​Tips on how to Tighten IST Table Rotation Motor Chain

​Tips on how to Tighten IST Table Rotation Motor Chain

Posted by " Jay the Table Guy" on Aug 26th 2024

Tips on how to Tighten IST Table Rotation Motor Chain

Troubleshooting =

1. Motor will not run -Remove Motor from table and apply 120V to motor to test. If Motor runs see below

2.If The motor is running and rollers feel jerky on patients back ,Its the chain is too loose.

SEE Tips on how to Tighten IST Table Rotation Motor Chain(Below)

3.The rotation Motor runs Intermittently , it is going to be a wire or wires in the wire harness 

Tips on how to Tighten IST Table Rotation Motor Chain

1. UNPLUG TABLE!!!!

2. Remove the upper wooden top from the table and sit to the side.

3. Loosen the 4 Bolts & nuts on Rotation Motor Bracket . (Loosen ONly it just need to slide slightly)

4. Pry with a BIG screw driver Between motor hole and bracket .

5.Keep Chain Tight and Tighten the 4 Bolts & nuts on Rotation Motor Bracket

5. Do NOT overtighten. You want the chain to be able to move slightly when squeezed with 2 fingers.

10. Plug in table, TURN OFF ALL MOTORS & VIBRATION .

11.Turn on  Timer then Rotation Motor ONLY, and test motor for noise. If the chain is noisy or pops when turning, then you have the chain tension to tight, adjust as needed.

11. Turn on all motors, watch and test again.

12. Install wooden top and you are ready for patients..

Rotation Motor Parts 

https://chiropractictableparts.com/search.php?sear...